                                       INSTRUMENT CONTROL PARAMETERS:    TW-GCMS
                                       -----------------------------------------

   C:\MSDCHEM\1\METHODS\TARGETED\HEIDI_FAMES.M
      Fri Sep 09 17:03:51 2016

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  CTC PAL ALS
Mass Spectrometer :  Enabled

============================================================================
                                CTCPAL METHOD
============================================================================
                    Injection Volume: 1.00 ul
                        Syringe Size: 10ul
                          Cycle File: GC-Inj

                                CYCLE DETAILS
                               ---------------
                     Air Volume (l): 0
         Pre Clean with Solvent 1 (): 1
         Pre Clean with Solvent 2 (): 1
            Pre Clean with Sample (): 0
                Filling Speed (l/s): 1
                  Filling Strokes (): 4
                           Inject to: GC Inj1
              Injection Speed (l/s): 100
               Pre Inject Delay (ms): 100
              Post Inject Delay (ms): 100
        Post Clean with Solvent 1 (): 1
        Post Clean with Solvent 2 (): 1

 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           0.5 min
Max Temperature                              325 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    60 C for 2 min
    then 10 C/min to 300 C for 5 min
Run Time                                     31 min




Sample Overlap
Sample overlap is not enabled



Front SS Inlet He
Mode                                         Splitless
Heater                                       On    250 C
Pressure                                     On    4.3986 psi
Total Flow                                   On    53.387 mL/min
Septum Purge Flow                            On    2 mL/min
Gas Saver                                    On    20 mL/min After 2 min
Purge Flow to Split Vent                     50 mL/min at 2 min


Thermal Aux 2 {MSD Transfer Line}
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     31 min


Column #1
Agilent HP-5msHP-5MS Part number 19091S
350 C: 15 m x 250 m x 0.25 m
In: Front SS Inlet He
Out: Vacuum

(Initial)                                    60 C
Pressure                                     4.3986 psi
Flow                                         1.3872 mL/min
Average Velocity                             61 cm/sec
Holdup Time                                  0.40984 min
Flow Program                                 Off
    1.3872 mL/min for 52 min
Run Time                                     31 min



Aux EPC 1 N2
Pressure Program                             Off
    10 psi for 0 min
Run Time                                     31 min

Aux EPC 2 N2
Pressure Program                             Off
    10 psi for 0 min
Run Time                                     31 min

Aux EPC 3 N2
Pressure Program                             Off
    10 psi for 0 min
Run Time                                     31 min


Signals
 Test Plot                                   Save Off
                                             50 Hz
 Test Plot                                   Save Off
                                             50 Hz
 Test Plot                                   Save Off
                                             50 Hz
 Test Plot                                   Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 6.00 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.00
Resulting EM Voltage     : 1329

[Scan Parameters]

Low Mass                 : 40.0
High Mass                : 600.0
Threshold                : 50
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 40.0
Plot 2 high mass         : 610.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US80838766
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      34.814
 IONFOCUS    :      90.157
 ENTRANCE_LE :      16.000
 EMVOLTS     :    1400.000
                               Actual EMV  :    1329.41
                               GAIN FACTOR :       0.96
 AMUGAIN     :    2046.000
 AMUOFFSET   :     120.250
 FILAMENT    :       2.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      18.322
 MASSGAIN    :    -456.000   
 MASSOFFSET  :     -40.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                                          END OF INSTRUMENT CONTROL PARAMETERS
                                          ------------------------------------